|
A. On
Cured Composite:
1. Contour composite.
2. Apply acid etchant,
such as UNI-ETCH®* w/BAC, TO COMPOSITE AND ADJACENT
TOOTH STRUCTURE for 15 seconds.
3. Rinse and dry.
4. Dispense BisCover
into a mixing well or apply directly using the syringe
and an applicator tip.
NOTE: The Viscosity Modifier may be added to
BisCover if a lower viscosity is desired. Do
not exceed a 1:1 ratio. Briskly mix for 5 seconds and
apply immediately to maintain the low viscosity throughout
the application. Allow 15 seconds before light curing
to ensure that the solvent has fully evaporated.
Do not apply unmixed Viscosity Modifier directly in
the mouth.
5. Apply one thin
coat of BisCover on cured composite. If needed,
air-thin BisCover by blowing a gentle stream
of air over this layer to ensure an even distribution.
CAUTION:
Upon curing, BisCover may produce a brief exothermic
reaction which is minimized by applying in a thin layer.
Do not cure on soft tissue.
6. Using a halogen
curing unit with a minimum output of 500mW/cm²
(7.5J/cm²), light cure for 30 seconds at close
range (0-2mm).
NOTE: If the light intensity is lower and/or
curing occurs at a distance (10mm) (both commonly occur
clinically), light cure for greater than 30 seconds.
(See chart in "What is EOP?" section.)
BisCover requires a conventional halogen curing
light in order to initiate polymerization. Insufficient curing
will leave an air-inhibited layer on the surface of
BisCover.
7. If a second coat
is required, repeat steps 4-6.
B. On
Uncured Composite:
1. Contour uncured composite to final anatomy.
2. Dispense and
apply one thin coat of BisCover on uncured composite
using a brush applicator. If needed, air-thin BisCover
by blowing a gentle stream of air over this layer to
ensure an even distribution.
NOTE: Viscosity
Modifier should not be used when applying BisCover
over uncured composite.
Caution:
Upon curing, BisCover may produce a brief exothermic
reaction which is minimized by applying in a thin layer.
Do not cure on soft tissue.
3. Light-cure according
to the composite manufacturer's recommendation.
OR Using a halogen curing unit with a minimum
output of 500mW/cm² (7.5J/cm²), light-cure
for 30 seconds at close range (0-2mm).
NOTE: If the light intensity is lower and/or
curing occurs at a distance (10mm) (both commonly occur
clinically), light-cure for greater than 30 seconds.
(See chart in "What is EOP?" section.)
BisCover requires a conventional halogen curing
light in order to initiate polymerization. Insufficient
curing will leave an air-inhibited layer on the surface
of BisCover.
4. If a second coat
is desired, repeat steps 2-3. If desired, the Viscosity
Modifier may be used in the second coat. See "NOTE"
for Viscosity Modifier in
section A step 4.
5. If occlusal adjustment
is required, reapply BisCover following instructions
for cured composite (section A).
C. On Old/Previously-Placed
Composites or for Reapplication:
1. Pumice, rinse
and dry the restoration.
2. Sandblast surface
or roughen with a fine diamond, rinse and dry.
3. Apply acid etchant,
such as UNI-ETCH w/BAC, for 15 sec.
4. Rinse with water (approx. 10-20 seconds) and dry thoroughly.
5. Dispense BisCover
into a mixing well or apply directly using the syringe
and applicator tip.
NOTE: The Viscosity Modifier may be added to
BisCover if a lower viscosity is desired. Do
not exceed a 1:1 ratio. Briskly mix for 5 seconds and
apply immediately to maintain the low viscosity throughout
the application. Allow 15 seconds before light curing
to ensure that the solvent has fully evaporated. Do
not apply unmixed Viscosity Modifier directly in the
mouth.
6. Apply one thin
coat of BisCover on cured composite. If needed,
air-thin BisCover by blowing a gentle stream
of air over this layer to ensure an even distribution.
CAUTION: Upon curing, BisCover may produce
a brief exothermic reaction which is minimized by applying
in a thin layer. Do not cure on soft tissue.
7. Using a halogen curing unit with a minimum output
of 500mW/cm² (7.5J/cm²), light cure for 30
seconds at close range (0-2mm).
NOTE: If the light intensity is lower and/or
curing occurs at a distance (10mm) (both commonly occur
clinically), light cure for greater than 30 seconds.
(See chart in "What is EOP?" section.)
BisCover requires a conventional halogen curing light to initiate
polymerization. Insufficient curing will leave an air-inhibited
layer on the surface of BisCover.
8. If a second coat
is desired, repeat steps 5-7.
D. Provisionals
and Appliances (Based on Acrylic Resins):
NOTE: For
large areas, the use of the Viscosity Modifier with
BisCover will allow it to spread more easily
and produce a thinner and more even film thickness.
1. Contour the provisional restoration.
2. Rinse and dry.
(Etchant is not required.)
3. Dispense BisCover
into a mixing well.
NOTE: The Viscosity Modifier may be added to
BisCover if a lower viscosity is desired. The
ratio of BisCover to Viscosity Modifier should
be 1:1. Briskly mix for 5 seconds and apply immediately
to maintain the low viscosity throughout the application.
Allow 15 seconds before light curing to ensure that
the solvent has fully evaporated.
4. Apply one thin
coat of BisCover. If needed, air-thin BisCover
by blowing a gentle stream of air over this layer to
ensure an even distribution.
5. Using a halogen
curing unit with a minimum output of 500mW/cm²
(7.5J/cm²), light cure each surface for 30 seconds
at close range (0-2mm).
NOTE: If the light intensity is lower and/or
curing occurs at a distance (10mm) (both commonly occur
clinically), light cure for greater than 30 seconds.
(See chart in "What is EOP?" section.)
BisCover requires a conventional halogen curing
light to initiate polymerization. Insufficient curing
will leave an air-inhibited layer on the surface of BisCover.
E. In
Orthodontic Applications:
Prior to Bracket
Placement:
1. Clean the surface to receive the BisCover/bracket
with a slurry of pumice/water.
2. Etch the surface
to receive the BisCover with 32% phosphoric acid, such
as UNI-ETCH w/BAC for 30 seconds (120 seconds for primary
teeth).
3. Rinse with water
(approx. 10-20 seconds) and dry thoroughly. The etched
area will appear frosty white. If not, re-etch for an
additional 20 seconds.
4. Dispense BisCover into a mixing well.
5. With a brush,
apply a thin layer of BisCover to the etched
enamel.
CAUTION: Upon curing, BisCover may produce
a brief exothermic reaction which is minimized by applying
in a thin layer. Do not cure on soft tissue.
6. Using a halogen
curing unit with a minimum output of 500mW/cm²
(7.5J/cm²), light cure the BisCover for
30 seconds per tooth at close range (0-2mm).
NOTE: If the light intensity is lower and/or
curing occurs at a distance (10mm) (both commonly occur
clinically), light cure for greater than 30 seconds.
(See chart in "What is EOP?" section.)
7. Apply desired
bracket cement (self-cured, light-cured or dual-cured)
to the bracket pad, and place bracket on tooth in the
usual manner. (No additional bonding resin is required.)
If Brackets have
already been placed:
1. Follow same instructions as in section E, "Prior
to Bracket Placement" excluding step 7.
| EOP
of BisCover is 7.5 Joules/cm2 or 7,500
mJ/cm2 |
What is EOP?
Energy
for Optimization of Polymerization is
the minimum amount of energy required to achieve clinically
acceptable polymerization. For BisCover, EOP
assumes a layer of 5-40µm. EOP assumes the tip
of the curing light is 1-2mm away from the surface to
be cured. Although this is easily achieved in the laboratory,
it is less likely to occur clinically, where the tip
is typically held at a distance (i.e. 10mm). EOP also
differs with the shade of the material beneath BisCover.
EOP was determined
by using a Quartz Tungsten Halogen curing unit (VIP,
BISCO, Schaumburg, IL) calibrated with a radiometer.
The effective wavelength for the curing light is 400-500nm.
The degree of polymerization of cured BisCover
is then measured spectroscopically. EOP is then calculated
as the energy required (Joules) to reach clinically
acceptable polymerization.
NOTE: BisCover
absorbs light in the range of 380-440nm. Therefore,
most LED curing units will not cure BisCover.
A known exception is the Ultra-Lume
LED 5 (Ultradent, Jordan, UT). To cure BisCover
with the Ultra-Lume LED 5, light cure for 20 sec. at
close range ( 5mm) or 40 sec. at a distance (10mm).
Plasma-arc (PAC)
light-curing units, such as the Q-Lux Plasma 100, require
10-15 seconds to cure BisCover.
Calculating Curing
Time for BisCover:
Energy delivered to the surface of BisCover depends
on the power density of the light, spectral output of
the light, exposure time and the distance between the
light probe and the BisCover coated surface. As mentioned,
the EOP of BisCover is 7,500mJ/cm². The power density
of a halogen curing light must first be determined using
a radiometer. For example, if the light output (PD,
power density) is 500mW/cm², then the calculation
is EOP/PD= 7,500/500= 15 seconds.
Although a distance
of 1-2mm from the material is easily achieved in the
laboratory, it is less likely to occur clinically where
the tip is typically held at a distance (i.e. 10mm).
The following chart describes the time requirements
needed for curing vs. distance from the light guide
at various Power outputs.
Power
Output of curing unit, mW/cm² |
Time,
sec. @ 0-2mm
(optimum clinical distance)
|
Time,
sec. @ 10mm
(typical clinical distance) |
750
|
10
|
20 |
500
|
15
|
30 |
375
|
20
|
>40 |
250
|
30
|
>60 |
Based on BisCover's
EOP of 7,500mJ/cm²
Using a Light
Box for Curing:
BisCover
can also be cured in commercially available light curing
boxes such as the Triad® 2000 (Dentsply,
York, PA). The preceding chart will help you determine
curing time. For most light boxes the power output will
be between 7.5-75mW. Refer to the chart as a guide.
Power
Output of curing unit, mW/cm² |
Time,
sec. @ 10mm
|
75
|
>200
(3.3 min) |
37.5
|
>400
(6.7 min) |
18.75
|
>800
(13.3 min) |
7.5
|
>2000
(33.3 min) |
Based on BisCovers
EOP of 7,500mJ/cm²
Please note that
BisCover cannot be overcured. Therefore, the BisCover coated appliance can be placed back into the processing
unit if it is determined that it has not been fully
cured.
Troubleshooting:
If cured BisCover
is tacky there are 4 possible reasons:
1. The light source
is less than the required amount. Bulb output decreases
with age. Use a radiometer to determine actual output.
Even though a set output has been dialed, many curing
units require calibration. Do not rely on dialed-in
values, unless they have been calibrated by using a
radiometer.
2. The light tip is too far away from the surface. Increase
curing time or increase output (PD) to obtain delivery
of 7,500mJ/cm² to the BisCover coated surface.
3. When coating uncured composite, BisCover layer
may have become contaminated with resin components from
the composite. In this case, add a second coat of BisCover
and repeat the curing process.
NOTE: When using BisCover over uncured
composite DO NOT use Viscosity Modifier. The second
coat of BisCover in this technique may be used
with Viscosity Modifier.
4. Light does not emit energy at the required wavelength
of 380-440nm.
STORAGE: Store at room temperature (22°C/72°F).
Refer to individual component labels for specific expiration
dates.
PRECAUTIONS: Unpolymerized resins may cause skin sensitization in
susceptible persons. In case of contact with the skin,
wash thoroughly with soap and water. Do not cure on
soft tissue.
WARRANTY: BISCO, Inc. recognizes its responsibility to replace
products if proven to be defective. BISCO, Inc. does
not accept liability for any damages or loss, either
direct or consequential, stemming from the use of or
inability to use the products as described. Before using,
it is the responsibility of the user to determine the
suitability of the product for its intended use. The
user assumes all risk and liability in connection therewith.
CAUTION: U.S. Federal law restricts this device to sale by or
on the order of a dental professional.
* BISCOVER is a trademark of BISCO, Inc.
UNI-ETCH and ONE-STEP are registered trademarks of BISCO,
Inc.
BISCO, Inc.
1100 W. Irving Park Rd.
Schaumburg, IL 60193
1-(800)-BIS-DENT
847-534-6000
www.bisco.com
CE ADMINISTRATOR
Bisico France -
Mr. Fred Picavet
L'Opera- BP. 60
13680 Lançon De Provence
France
Phone: 33-490-42-9292
Fax: 33-490-42-9261
IN-117R4
Rev. 10/07
U.S. Patent: 7,081,485
|