|
A. On Cured Composite:
1. Contour composite.
2. Apply acid
etchant, such as UNI-ETCH®* w/BAC, TO COMPOSITE AND ADJACENT
TOOTH STRUCTURE for 15 seconds.
3. Rinse and
dry.
4. Dispense
BisCover into a mixing well or apply directly using the
syringe and an applicator tip.
NOTE: The Viscosity Modifier may be added to BisCover
if a lower viscosity is desired. Do not exceed a 1:1 ratio. Briskly
mix for 5 seconds and apply immediately to maintain the low viscosity
throughout the application. Allow 15 seconds before light curing
to ensure that the solvent has fully evaporated. Do not apply
unmixed Viscosity Modifier directly in the mouth.
5. Apply one
thin coat of BisCover on cured composite. If needed, air-thin
BisCover by blowing a gentle stream of air over this layer
to ensure an even distribution.
CAUTION:
Upon curing, BisCover may produce a brief exothermic reaction
which is minimized by applying in a thin layer. Do not cure
on soft tissue.
6. Using a
halogen curing unit with a minimum output of 500mW/cm² (7.5J/cm²),
light cure for 30 seconds at close range (0-2mm).
NOTE: If the light intensity is lower and/or curing occurs
at a distance (10mm) (both commonly occur clinically), light cure
for greater than 30 seconds. (See chart in "What is EOP?"
section.)
BisCover requires a conventional halogen curing light in order to
initiate polymerization. Insufficient curing will leave an air-inhibited
layer on the surface of BisCover.
7. If a second
coat is required, repeat steps 4-6.
B. On
Uncured Composite:
1. Contour uncured composite to final anatomy.
2. Dispense
and apply one thin coat of BisCover on uncured composite
using a brush applicator. If needed, air-thin BisCover
by blowing a gentle stream of air over this layer to ensure an
even distribution.
NOTE: Viscosity
Modifier should not be used when applying BisCover over
uncured composite.
CAUTION:
Upon curing, BisCover may produce a brief exothermic reaction
which is minimized by applying in a thin layer. Do not cure
on soft tissue.
3. Light cure
according to the composite manufacturer's recommendation.
OR Using a halogen curing unit with a minimum output of
500mW/cm² (7.5J/cm²), light cure for 30 seconds at close
range (0-2mm).
NOTE: If the light intensity is lower and/or curing occurs
at a distance (10mm) (both commonly occur clinically), light cure
for greater than 30 seconds. (See chart in "What is EOP?"
section.)
BisCover requires a conventional halogen curing light in
order to initiate polymerization. Insufficient curing will leave
an air-inhibited layer on the surface of BisCover.
4. If a second
coat is desired, repeat steps 2-3. If desired, the Viscosity
Modifier may be used in the second coat. See "NOTE" for Viscosity
Modifier in section A step 4.
5. If occlusal
adjustment is required, reapply BisCover following instructions
for cured composite (section A)
C. On
Old/Previously-Placed Composites or for Reapplication:
1. Pumice,
rinse and dry the restoration.
2. Sandblast
surface or roughen with a fine diamond, rinse and dry.
3. Apply acid
etchant, such as UNI-ETCH w/BAC, for 15 sec.
4. Rinse with water (approx. 10-20 seconds) and dry thoroughly.
5. Dispense
BisCover into a mixing well or apply directly using the
syringe and applicator tip.
NOTE: The Viscosity Modifier may be added to BisCover
if a lower viscosity is desired. Do not exceed a 1:1 ratio. Briskly
mix for 5 seconds and apply immediately to maintain the low viscosity
throughout the application. Allow 15 seconds before light curing
to ensure that the solvent has fully evaporated. Do not apply
unmixed Viscosity Modifier directly in the mouth.
6. Apply one
thin coat of BisCover on cured composite. If needed, air-thin
BisCover by blowing a gentle stream of air over this layer
to ensure an even distribution.
CAUTION: Upon curing, BisCover may produce a brief
exothermic reaction which is minimized by applying in a thin layer.
Do not cure on soft tissue.
7. Using a halogen curing unit with a minimum output of 500mW/cm²
(7.5J/cm²), light cure for 30 seconds at close range (0-2mm).
NOTE: If the light intensity is lower and/or curing occurs
at a distance (10mm) (both commonly occur clinically), light cure
for greater than 30 seconds. (See chart in "What is EOP?"
section.)
BisCover requires a conventional halogen curing
light to initiate polymerization. Insufficient curing will leave
an air-inhibited layer on the surface of BisCover.
8. If a second
coat is desired, repeat steps 5-7.
D. Provisionals
and Appliances (Based on Acrylic Resins):
NOTE:
For large areas, the use of the Viscosity Modifier with BisCover
will allow it to spread more easily and produce a thinner and
more even film thickness.
1. Contour the provisional restoration.
2. Rinse and
dry. (Etchant is not required.)
3. Dispense
BisCover into a mixing well.
NOTE: The Viscosity Modifier may be added to BisCover
if a lower viscosity is desired. The ratio of BisCover
to Viscosity Modifier should be 1:1. Briskly mix for 5 seconds
and apply immediately to maintain the low viscosity throughout
the application. Allow 15 seconds before light curing to ensure
that the solvent has fully evaporated.
4. Apply one
thin coat of BisCover. If needed, air-thin BisCover
by blowing a gentle stream of air over this layer to ensure an
even distribution.
5. Using a
halogen curing unit with a minimum output of 500mW/cm² (7.5J/cm²),
light cure each surface for 30 seconds at close range (0-2mm).
NOTE: If the light intensity is lower and/or curing occurs
at a distance (10mm) (both commonly occur clinically), light cure
for greater than 30 seconds. (See chart in "What is EOP?"
section.)
BisCover requires a conventional halogen curing light to
initiate polymerization. Insufficient curing will leave an air-inhibited
layer on the surface of BisCover.
E. In
Orthodontic Applications:
Prior to
Bracket Placement:
1. Clean the surface to receive the BisCover/bracket with a slurry
of pumice/water.
2. Etch the
surface to receive the BisCover with 32% phosphoric acid, such
as UNI-ETCH for 30 seconds (120 seconds for primary teeth).
3. Rinse with
water (approx. 10-20 seconds) and dry thoroughly. The etched area
will appear frosty white. If not, re-etch for an additional 20
seconds.
4. Dispense
BisCover into a mixing well.
5. With a
brush, apply a thin layer of BisCover to the etched enamel.
CAUTION: Upon curing, BisCover may produce a brief
exothermic reaction which is minimized by applying in a thin layer.
Do not cure on soft tissue.
6. Using a
halogen curing unit with a minimum output of 500mW/cm² (7.5J/cm²),
light cure the BisCover for 30 seconds per tooth at close
range (0-2mm).
NOTE: If the light intensity is lower and/or curing occurs
at a distance (10mm) (both commonly occur clinically), light cure
for greater than 30 seconds. (See chart in "What is EOP?"
section.)
7. Apply desired
bracket cement (self-cured, light-cured or dual-cured) to the
bracket pad, and place bracket on tooth in the usual manner. (No
additional bonding resin is required.)
If Brackets
have already been placed:
1. Follow same instructions as in section E, "Prior to Bracket
Placement" excluding step 7.
| EOP
of BisCover is 7.5 Joules/cm2 or 7,500 mJ/cm2 |
What is
EOP?
Energy
for Optimization of Polymerization is the minimum
amount of energy required to achieve clinically acceptable polymerization.
For BisCover, EOP assumes a layer of 5-40µm. EOP
assumes the tip of the curing light is 1-2mm away from the surface
to be cured. Although this is easily achieved in the laboratory,
it is less likely to occur clinically, where the tip is typically
held at a distance (i.e. 10mm). EOP also differs with the shade
of the material beneath BisCover.
EOP was determined
by using a Quartz Tungsten Halogen curing unit (VIP, BISCO, Schaumburg,
IL) calibrated with a radiometer. The effective wavelength for
the curing light is 400-500nm. The degree of polymerization of
cured BisCover is then measured spectroscopically. EOP
is then calculated as the energy required (Joules) to reach clinically
acceptable polymerization.
NOTE: BisCover
absorbs light in the range of 380-440nm. Therefore, most LED
curing units will not cure BisCover. A known exception
is the Ultra-Lume LED 5 (Ultradent, Jordan, UT).
To cure BisCover with the Ultra-Lume LED 5, light cure
for 20 sec. at close range ( 5mm) or 40 sec. at a distance (10mm).
Plasma-arc
(PAC) light curing units, such as the Q-Lux Plasma 100, require
10-15 seconds to cure BisCover.
Calculating
Curing Time for BisCover:
Energy delivered to the surface of BisCover depends on the power
density of the light, spectral output of the light, exposure time
and the distance between the light probe and the BisCover coated
surface. As mentioned, the EOP of BisCover is 7,500mJ/cm².
The power density of a halogen curing light must first be determined
using a radiometer. For example, if the light output (PD, power
density) is 500mW/cm², then the calculation is EOP/PD= 7,500/500=
15 seconds.
Although a
distance of 1-2mm from the material is easily achieved in the
laboratory, it is less likely to occur clinically where the tip
is typically held at a distance (i.e. 10mm). The following chart
describes the time requirements needed for curing vs. distance
from the light guide at various Power outputs.
Power
Output of curing unit, mW/cm² |
Time,
sec. @ 0-2mm
(optimum clinical distance)
|
Time,
sec. @ 10mm
(typical clinical distance) |
750
|
10
|
20 |
500
|
15
|
30 |
375
|
20
|
>40 |
250
|
30
|
>60 |
Based on BisCover's
EOP of 7,500mJ/cm²
Using a
Light Box for Curing:
BisCover
can also be cured in commercially available light curing boxes
such as the Triad® 2000 (Dentsply, York, PA). The
preceding chart will help you determine curing time. For most
light boxes the power output will be between 7.5-75mW. Refer to
the chart as a guide.
Power
Output of curing unit, mW/cm² |
Time,
sec. @ 10mm
(typical clinical distance) |
75
|
>200
(3.3 min) |
37.5
|
>400
(6.7 min) |
18.75
|
>800
(13.3 min) |
7.5
|
>2000
(33.3 min) |
Based on BisCovers EOP of 7,500mJ/cm²
Please note
that BisCover cannot be overcured. Therefore, the BisCover coated
appliance can be placed back into the processing unit if it is
determined that it has not been fully cured.
Troubleshooting:
If cured BisCover
is tacky there are 4 possible reasons:
1. The light
source is less than the required amount. Bulb output decreases
with age. Use a radiometer to determine actual output. Even though
a set output has been dialed, many curing units require calibration.
Do not rely on dialed-in values, unless they have been calibrated
by using a radiometer.
2. The light tip is too far away from the surface. Increase curing
time or increase output (PD) to obtain delivery of 7,500mJ/cm²
to the BisCover coated surface.
3. When coating uncured composite, BisCover layer may have
become contaminated with resin components from the composite.
In this case, add a second coat of BisCover and repeat
the curing process.
NOTE: When using BisCover over uncured composite
DO NOT use Viscosity Modifier. The second coat of BisCover
in this technique may be used with Viscosity Modifier.
4. Light does not emit energy at the required wavelength of 380-440nm.
STORAGE: Store at room temperature (22°C/72°F).
Refer to individual component labels for specific expiration dates.
PRECAUTIONS:
Unpolymerized resins may cause skin sensitization in susceptible
persons. In case of contact with the skin, wash thoroughly with
soap and water. Do not cure on soft tissue.
WARRANTY:
BISCO, Inc. recognizes its responsibility to replace products
if proven to be defective. BISCO, Inc. does not accept liability
for any damages or loss, either direct or consequential, stemming
from the use of or inability to use the products as described.
Before using, it is the responsibility of the user to determine
the suitability of the product for its intended use. The user
assumes all risk and liability in connection therewith.
CAUTION:
U.S. Federal law restricts this device to sale by or on the order
of a dental professional.
* BisCover
is a trademark of BISCO, Inc.
UNI-ETCH is a registered trademark of BISCO, Inc.
BISCO, Inc.
1100 W. Irving Park Rd.
Schaumburg, IL 60193
1-(800)-BIS-DENT
847-534-6000
www.bisco.com
CE ADMINISTRATOR
Bisico France -
Mr. Fred Picavet
L'Opera- BP. 60
13680 Lancon De Provence
France
Phone: 33-490-42-9292
Fax: 33-490-42-9261
IN-117R4
Rev. 10/07
U.S. Patent: 7,081,485
|